ACMAL spring break schedule

I will be out next week but I’ll mostly be in town. I could come in if needed. Aleister will be here working working 2 hours each day for coating samples and helping users. He plans to be here at midday.

The labs will be open all week between 8 – 5 and restricted users will work then. Unrestricted users have 24 hour access. Newly trained users, please work while Aleister is at work in case you have trouble.

Call Aleister or me if you have trouble. Aleister 520-576-3557. Owen 906-369-1875.

Thanks, Owen

IMPORTANT Saturating ESEM filament

We have made changes to the process of saturating the ESEM filament.  Now we use Scan>Horizontal Line and monitor that line as the Filament slider is changed.  Saturation occurs when the horizontal line stops increasing when the slide is increased.  The point when it stops increasing is saturation, so you may overshoot and have to slide it back slightly.  When you complete saturation return to Scan>Full Screen to stop the line.  Click Filament Limit as you usually would.  Any of you who learned to saturate the filament on the 6400 SEM will recognize this technique.

I want to show you how to do this so let me know when you’d want to see it done.  It is really easy and there is no ambiguity when setting saturation like using the “football”.  We will be making changes to the standard operating procedure.

This a a precise technique and the filament will last longer.

ESEM back online now

The ESEM is back online and ready to use.

I’m not certain what went wrong but I’m suspicious of the AZtec scan grabber. The AZtec software acquires an ESEM image by taking control of the ESEM scan hardware. When the scan is complete it should release the ESEM hardware. I have seen a few times that it doesn’t release. If you have trouble seeing a scanning ESEM image after acquiring with AZtec try to shut down the AZtec software and see if the scan returns. If it still down’t return contact me and I’ll take a look at it.

Liz Miller’s schedule

Liz Miller will be out of town for 6 weeks beginning next week. Aleister and I will try to pick up the work she was doing during her absence. One area that he and I cannot assume is EBSD. We have asked someone to help us with that but the response will be slowed. So, plan your EBSD work ahead of time – don’t wait until the last minute or your work will be delayed.

Liz will return to work on April 6th.

ESEM EDS detector alarm

EDS detectors works like a thermos bottle that slow liquid nitrogen evaporation. When the vacuum in the detector deteriorates the liquid nitrogen that cools the detector evaporates quicker. That is happening in the ESEM detector. We refill the detector twice each week but depending on the timing you might hear an alarm signaling that the liquid nitrogen is low. The alarm is a beep signal every few seconds. Here is the important part of this message – if you hear the alarm let me, Aleister or Liz know so we can refill it.

I will have the vacuum replenished in the summer when ESEM use drops.

Thanks, Owen

Analytical Electron Microscopy at Nanoscale Feb. 13

Pinaki Mukherjee
Pinaki Mukherjee

ME-EM Graduate Seminar Speaker Series

proudly presents:

Pinaki Mukherjee, PhD

Michigan Technological University

Abstract: This talk presents an overview of state-of-the-art capabilities of the aberration-corrected scanning transmission electron microscope (ac-STEM) at ACMAL, Michigan Tech. The FEI Titan Themis microscope we have here is one of a kind in the whole nation in terms of capabilities. These capabilities have been developed in last two years and most of them are already available for users. We have a wide range of imaging and spectroscopic techniques that enables a user to identify elements at atomic scale (~ 0.1 nm). READ MORE

Thursday, February 13 at 4:00 p.m.

Electrical Energy Resources Center (EERC), 103
1400 Townsend Drive, Houghton, MI 49931