February 12, 2024
Update on the FESEM,
- The microscope now has a clean version of the operating system.
- This week, MTU IT will be setting up a system to allow for retrieval of the data.
As of today, do NOT put any USB into the FESEM.
February 5, 2024
Just an update. Unfortunately the FESEM is still down. We need to rebuild the hard drive for this computer. HITACHI will be on campus this week to resolve the issue.
January 26, 2024
All FESEM users need to be aware that there was a virus on the FESM that had been transmitted by the USB drive. HITACHI was on campus this week and confirmed the infection. We are working to get a clean harddrive in the instrument. Until then the FESEM will be unavailable.
Additionally, anyone who has used the USB to transfer files may have malware on their USB and computer. Please work with MTU IT if you have any concerns.
Once the FESEM is back online we will be establishing a new protocol for transferring files. More information will be available soon.